By Bruce T. Stanley

Summary: The two primary ways of producing ozone for water treatment are through corona discharge or electrolytically. This article discusses advantages of the latter in contrast with other disinfection technologies.


Over the years, manufacturers have grown increasingly aware of the importance of water purity and its effect on the quality of the final product. One of the largest problems that confronts pure water system operators is bacterial recontamination shortly after the water purification equipment has done its job. There are several recommended methods of either preventing or removing such contamination but most have inherent disadvantages. The most innovative and effective method adopted by leading pharmaceutical, biotech, cosmetics and semiconductor manufacturers, involves ozonation of the water in the system and reducing this prior to the first point-of-use with ultraviolet (UV) irradiation.

An efficient way of producing ozone is with an electrolytic ozone generator that actually produces ozone from the water being treated. Ozone, and particularly electrolytically generated ozone, is ideally suited for pure water loops because only low concentrations are necessary to sanitize the system and there are no objectionable by-products or residue after the ozone has decomposed back to oxygen.

Controlling microbial contamination
Irrespective of how well a plant has been designed and constructed, when no special measures have been taken, it’s virtually impossible to avoid microbial contamination in a pure water network. This is especially the case when a system isn’t continually replenished with fresh makeup water such as overnight or over the weekend when production is shut down and the water stands in a tank or circulates in a closed loop.

Only for specific applications do regulatory bodies require certain criteria are met and request operators of such plants to carry out validation tests. In all other cases, operators are left to their own means and follow accepted guidelines to establish and maintain the integrity of their pure water systems.

Weighing methods
There are several standard methods used by operators to disinfect their pure water systems, however, each has disadvantages listed below that have a bearing on the water quality:

  1. Shock disinfection—with chemicals such as peroxide or hyperchlorites.
    a) Interruption of operation
    b) Work intensive
    c) Associated chemical problems
    d) Fluctuating water quality
    e) Safety concerns
  2. Shock sterilization—with steam.
    a) Interruption of operation
    b) Costly installation and service
    c) Uncontaminated steam necessary
    d) Fluctuating water quality
    e) Possible “cold spots,” i.e., areas of reduced disinfection
  3. UV irradiation.
    a) Risk of reduced disinfection
    b) Only localised effect
    c) Sterile filter beneficial
    d) Regular component replacement
  4. Sterile filtration.
    a) Bacterial growth is unaffected
    b) Regular replacement required
    c) Danger of bursting
    d) Expensive

Ozone advantages
Of the above mentioned standard methods shock disinfection with chemicals or shock sterilization with steam seem to produce the best results provided service interruptions and fluctuating water quality can be tolerated. An innovative and alternative method of disinfection of pure water loops, without any of the mentioned drawbacks, is by introducing ozone to the circulation flow. The advantages of using ozone are manifold:

  • There are no objectionable by-products or residues when water is disinfected with ozone. In the absence of oxidizable substances, ozone decomposes to form oxygen; as soon as oxidizable substances are present, traces of carbon dioxide will form. These substances do not pose a major problem in connection with the water quality.
  • Because the ozone does ultimately decay to form oxygen—the ozone molecule is only moderately stable and has a half fife time of something like 30 to 60 minutes at normal service conditions—there are no lasting problems with traces of disinfection chemicals.
  • Experience gained in the pharmaceutical industry has shown very low ozone concentrations in the magnitude of 0.1 to 0.2 mg/L are sufficient to keep germ counts below 1 colony-forming units (cfu) per 100 milliliters (ml). Figure 1 indicates microbial growth as a function of ozone concentration.
  • Disinfection with ozone is a continual process and, in most cases, can be regulated by simple means. If required, however, the ozone production rate can be controlled by the process parameters in order to avoid incorrect dosing and to ensure optimized efficiency.
  • Owing to ozone’s limited half-life, it has to be produced in-situ when and where it’s required. Although this means extra investment must be made for a minor additional amount of infrastructure in the form of an ozone generator it avoids the transportation of potentially dangerous chemicals or the high running costs of a steam sterilization unit.
  • The Electrochemical Oxidation Potential (EOP) of ozone is considerably higher than other disinfecting agents (see Table 1).

Ozone generation
The state-of-the-art method of producing ozone in this manner is by means of dielectric barrier discharge (DBD) using either dry air or dry oxygen as the feed gas. Synthetic ozone produced by these generators is ideal for drinking water disinfection waste water treatment, pulp bleaching, etc., but does have limitations when dealing with pure water systems. For example, fluctuations in water pH and resistivity can occur if feed gas preparation (oxygen) isn’t properly maintained.

A far more elegant way of producing ozone for this specific application without the inherent disadvantages associated with conventional processes is with an electrolytic ozone generator that splits water into its basic elements and then converts part of the liberated oxygen into ozone.

The main advantages of producing ozone with an electrolytic system are:

  1. There is no ionic contamination because the feed water is being dissociated using a solid hydrated ion exchange membrane.
  2. The process water being disinfected is the source of oxygen for the generation of ozone—consequently, no outside contamination is introduced into the system being treated.
  3. The ozone is dissolved in the process water as soon as it is formed, resulting in ozonation with minimal equipment.
  4. By operating the cell under pressure, relatively high ozone concentrations can be produced.

This electrolytic system is fundamentally different method of dissociating water due to the fact it uses a solid polymer membrane as the electrolyte instead of fluid. This feature in conjunction with suitably controlled intermediate anodic reaction makes it particularly suited for disinfecting pure water systems irrespective of their application. A basic cell is shown in Figure 2.

The membrane, which functions as both electrolyte and separator between anode and cathode, is contacted on both sides by the activated porous electrodes. The water fed to the anode side of the cell is dissociated at the interface between the anode and membrane as a result of the DC current being applied. To ensure that as much ozone as possible is produced, the anode must have an over potential above the decomposition and the ozone reaction potential, and the catalytic layer must inhibit the formation of diatomic oxygen and encourage the formation of ozone.

Integration
Most modem pure water networks, are constructed as closed loop systems in which the water is pumped through one or more circulation loops to different consumer points. Depending on the application of the process water, there will be differing amounts and types of equipment installed in the system. Figures 3, 4 and 5 show typical flow diagrams for three different uses.

For all applications, it was found best to install the ozone generation system in the loop and return just before it re-enters the storage tank. The by-pass flow for the electrolytic system is tapped off before the loop’s pressure retaining valve and reintroduced on the low pressure side (see Figure 6).

Ozone effect
Ozone is a very strong oxidising agent with powerful disinfecting properties. It’s bactericidal, viricidal, fungicidal and sporicidal—furthermore, it’s active against yeasts, parasites, etc. One of the important factors to be taken into consideration when calculating the ozone dose is the organic content of the water being disinfected. That’s because, if this is high, the ozone demand will be correspondingly high and the ozone also will be used to reduce the organic content rather than just destroying microorganisms. During periods when ozone is being used to oxidize both microorganisms and organics, the kill rate will be slow. Once the total organic carbon (TOC) has been reduced to a minimum, the kill rate will increase accordingly. It should be borne in mind, however, that because cell death is caused by lysis and catoplasmic dispersion, i.e., it ruptures the cell wall and attacks the RNA, this will continually supplement the organic content of the water.

In order to simulate the effect of electrolytically produced ozone on normal water contamination, a 1988 study2 carried out tests on a mixed culture of Staphylococcus albus, Pseudomonas diminuta, Flavobacterium devorans and Coryneybacterium pyrogenes which were seeded into a tank prior to ozonation. A count was taken for the total quantity of bacteria killed rather than the kill for each particular species (see Table 2).

All species of bacteria inoculated could be detected in the initial plates. It was noted that all of the species were removed (see also Figures 7&8).

A second test was performed using the same mixture of cultures but with a higher ozone production rate (see Table 3). Again, a count was taken for the total quantity of bacteria killed rather than the kill for each particular species.

The water used for the second test had a higher temperature than the water used for the first test. Although some cooling did take place during the test period, the lack of detected residual ozone can be considered a result of the higher temperature reducing the half-life. Ozone did start to build up after 80 minutes when the temperature had fallen to 28°C. Even though the residual ozone content was below the detectable limit, the bacteriological reduction was rapid and can be attributed to the synergistic effect of ozone and temperature (see Figures 9&10). During testing, it was noted that the Staphylococcus albus and the Pseudomonas diminuta appeared to have the greatest resistance to ozone, while the other species were reduced rapidly.

In addition to the tests on bacteria, the 1988 study also demonstrated that ozone was very effective against other microorganisms:

  • Virus:Ozone was proved to be a very powerful viricide, e.g., Poliovirus was inactivated in 2 minutes with an ozone concentration of 0.05-0.45 mg/L.
  • Cysts: Enteramoeba hystolytica were removed with an ozone concentration of 0.3 mg/L for 2.4 minutes.
  • Spores: Ten to 15 times more resistant to ozone than vegetative cells because of the extra protection afforded by the spore coat.
  • Fungi: Candida albicans and a Penicillium were successfully sterilised.
  • Parasites: Schistosoma mansoni were sterilised after 3 minutes.

Ozone destruction
Depending on the process or product in question, it’s sometimes necessary to remove the ozone from the water prior to use. There are several methods of removing or reducing ozone (activated carbon filters, catalytic conversion units, thermal destructors, UV, etc.) but, because of the nature of pure water and associated systems, only UV irradiation comes into question.

Normal practice is to install a UV unit in the piping system shortly before the first point-of-use. This is switched on shortly before starting the main process or commencing with production and then switched off when there’s no demand for a longer period of time such as overnight or over the weekend. The incorporation of a UV ozone destructor unit also offers operators the small added benefit gained from the fact that hydroxyl-radicals, with a high oxidation potential, are produced in the lamp unit.

Standard germicidal UV units—with low-pressure mercury lamps and a high UV-C output at the wavelength of 254 nanometers—are ideal for reducing ozone in pure water loops to below the measurable limits. When dimensioning such lamps, special attention must be given to the fact the UV dose delivered at the end of the lamp’s service fife, and at maximum water flow, is sufficient to achieve the required level of ozone reduction. Figure 11 shows the ozone reduction rate at various UV doses.

Conclusion
Experience gained from numerous installations in various branches of industry and institutions indicate the sanitisation of pure water systems with ozone is an easy and effective method not requiring specialised operator skills.

Because ozone disinfection is a continual process allowing very little chance of (re)contamination, many operators now rely solely on the ozone treatment and have dispensed with regular shock disinfection with either steam or other chemicals.

Acknowledgment
The author would like to thank Gaspar Lesznik for his assistance with this article. Lesznik has been with Ozonia North America and its predecessor companies for 29 years, during which he has been involved with all aspects of ozone application engineering, sales, marketing and product development.

References

  1. MacRae, R.J., et al., “Sterilisation of Water by Ozone,” Corporate Development Report, Fisons Plc, Ipswich, Suffolk, United Kingdom, 1988.
  2. Möll, F., “Ozonation of Pure Water used in the Manufacture of Medicines,” Proceedings of the CONCEPT Symposium, Fulda, Germany (Spirig Ltd), 1990.
  3. Klein, H-P, and B. Steinlin, “Electrolytic Ozone Generation and its Uses,” Corporate Report, Brown Boveri Co. Ltd, 1989.
  4. Kogelschatz, U., “Advanced Ozone Generation,” Proceedings of the 10th Brown Boveri Symposium on Process Technologies for water Treatment, Brown Boveri Research Center, 1987.
  5. Stucki, S., and H. Baumann, “Application of Electrochemical Ozone Generators in Ultrapure Water Systems,” Proceedings of the 10th Brown Boveri Symposium on Process Technologies for Water Treatment, Brown Boveri Research Center, Baden, Switzerland, 1987.

About the authors
Bruce Stanley has worked for Ozonia Ltd. and its predecessor ABB, in Duebendorf, Switzerland, for more than 30 years. Stanley has been involved in all aspects of ozone generation and use of the technology (corona and electrolytic) in the years he’s been with Ozonia. He was part of the original group that designed and developed the MEMBREL® process, discussed above using a hydrogen cell for ozone generation. Ozonia has branches and factories in Switzerland, the United States, Scotland, South Korea and Russia. It’s a leader in development, manufacture and supply of ozone generation and medium-pressure UV systems for environmentally responsible disinfection solutions.